In the daily use of the dew point osmometer 5600/5520, contamination of the thermocouple head can affect the accuracy of the sample test results, which often requires cleaning the thermocouple head. This video will show you how to properly clean the thermocouple head. Simple operation of cleaning the thermocouple head: Dew point osmometer: http:// For the repair of wescor series products, please call us! Medical Hospital Bed,Patient Bed,Electric Hospital Bed,Adjustable Hospital Bed Hebei Dingli Medical Apparatus and Instrument Co., LTD , https://www.dinglimed.com
1. Press the right side of the small cover on the top of the instrument and remove the cover.
2. Use a hexagonal screwdriver equipped with the instrument to remove the thermocouple head (Note: Do not touch the thermocouple wire under the thermocouple, as this may cause damage to the thermocouple)
3. Place the thermocouple head under the microscope with the silky part facing up. Cut one end of the swab into a pointed shape and use a pointed tip to gently scrape off dirt on or near the wire. (Note: Do not touch the thermocouple wire under the thermocouple, as this will cause the thermocouple to be damaged)
4. Drain the cleaning solution and soak it for a few minutes.
5. Wash the cleaning solution with deionized water and remove it.
6. Dry the surface of the thermocouple head with a blower.
7. Place the dry thermocouple head inside the instrument.
Special note: Please do the whole operation under the microscope.